VWR 444-7084 Manual De Instrucciones página 7

Tabla de contenido
Idiomas disponibles

Idiomas disponibles

CAUTION HOT indicator:
The caution hot indicator light warns that the temperature
of the air in the chamber is above 40°C (104°F). The
light will illuminate and remain lit when the temperature of the air in
the chamber reaches approximately 40°C (104°F). When the heat is
turned off, the caution hot indicator light will stay lit until the temperature
of the air in the chamber is less than 40°C (104°F).
3. Setting speed:
a. Press the up/down arrows below the speed display until you reach
the desired speed. When you release the button, the display will
blink off and then on indicating the new set speed has been
accepted.
b. Press the on/off button to start the shaking function. The indicator
light below the speed display will illuminate and blink until the
set-point is reached. Once the set-point is reached the light will
stop blinking and remain lit until shaking has ceased. The micro-
processor controlled ramping feature slowly increases speed until
the set-point is reached which helps to avoid splashing, and provides
excellent low end control.
c. Speed adjustments can be made without interrupting shaking by
using the up/down arrows below the speed display. After the change
has been made and you release the button, the display will blink
off and then on indicating the new set speed has been accepted.
d. To stop the shaking function, press the on/off button below the
speed display. The speed indicator light will turn off.
4. Setting time to zero (0:00) and continuous mode: Accumulated time.
a. Press and hold the on/off button below the time display. After
three (3) seconds the display will indicate the previous set time.
b. Simultaneously press both the up and the down arrows, the display
will indicate zero (0:00). The unit time is now set to zero (0:00)
minutes. Alternately, you can use the up/down arrows to get to
zero (0:00).
c. Press the on/off button below the time display. The display will
indicate accumulated time. The up/down arrows will become inactive.
To stop timer, press the on/off button again. IMPORTANT: This will
NOT interrupt the shaking function. Press the on/off button below
the speed display to interrupt the shaking function.
d. To reset, press and hold the on/off button below the time display.
After three (3) seconds the display will indicate the previous set
time, which was zero (0:00).
5. Setting timed mode: Programmed time.
a. Press the up/down arrows below the time display until you reach
the desired time.
b. Start this function by pressing the on/off button below the time
display. The unit will run for the selected time, the up/down arrows
will become inactive while the timer is running. The unit will stop
shaking when time display reaches zero (0:00). Four (4) audible
beeps will indicate the count down function is complete. The time
display will default back to the set time. To repeat for the same
time, simply press the on/off button again.
c. To interrupt an automatic timing cycle before it is completed, press
the on/off button below the time display. The display will flash off
and on to indicate the time function is on "hold". IMPORTANT: This
will NOT interrupt the shaking function. Press the on/off button
below the speed display to interrupt the shaking function. Restart
the timer by pressing the on/off button below the time display.
Unit will continue counting down to zero (0:00). When the display
reaches zero (0:00), you will hear the four (4) audible beeps that
indicate the count down function is complete and the shaking
function will cease.
6
Tabla de contenido
loading

Este manual también es adecuado para:

44-7085444-7086

Tabla de contenido