Accessing old measurement values

1.
1. Starting position.
2. Press the right or left selector button until Old appears on the display. Then press the measurement button.
The old values can now be scanned back and forward by pressing the selector buttons (right= older, left=recent).
3. Press the measurement button to exit the old values search. Old is on the display. Press either of the selector
buttons to access the other functions (00=measurement, End=turning OFF).
Changing the calibration table
for different species
When the display shows End (p. 4), press first measurement, keep it pressed down and immediately
press right selector button. Buttons have to be pressed down at the same time. Then the display shows
CAL (or software version). Toggle between these using selector button and choose CAL by pressing
the measurement button. After it shows the menu where to choose the calibration table:
do = dog /cat
ho = horse
P = not defined, reserved for future use.
Toggle between these using selector buttons and choose the needed calibration table by pressing
the measurement button. When measuring, the result display shows which calibration table is in use
(d, H or P). The default selection is d (dog/ cat).
8
2.
3.