[ I/1 ] AEx [ib] / IIB - 2003809
Specific conditions of use
1. The YCO16-Z Ex-Link Converter shall only be used with the PMA.Vision VIS1X...-....
2. The protective ground connectors on the PMA.Vision VIS1X...-... and the YCO16-Z Ex-Link Converter shall be connected
together using a potential equalization conductor with a cross-sectional area of at least 4 mm
3. The CAT6 cable used between the PMA.Vision VIS1X...-... and the YCO16-Z Ex-Link Converter shall be obtained from
Sartorius.
YCO17-Z Ex Link Converter
ANI / I / 1 / CD – 2003809
Specific conditions of use
1. The YCO16-Z Ex-Link Converter shall only be used with the PMA.Vision VIS1Y...-....
2. The protective ground connectors on the PMA.Vision VIS1Y...-... and the YCO16-Z Ex-Link Converter shall be connected
together using a potential equalization conductor with a cross-sectional area of at least 4 mm
3. The CAT6 cable used between the PMA.Vision VIS1Y...-... and the YCO16-Z Ex-Link Converter shall be
Equipment Ratings:
PMA.Vision VIS1X...-...
Intrinsically Safe for Class I, Division 1, Groups C, and D, hazardous (classified) locations in
accordance with drawing 2003809; and Intrinsically Safe for Class I, Zone 1, Group IIB hazardous
(classified) locations in accordance with drawing 2003809.
PMA.Vision VIS1Y...-...
Nonincendive for Class I, Division 2, Groups C, and D, hazardous (classified) locations in
accordance with drawing 2003809; and Intrinsically Safe for Class I, Zone 1, Group IIB hazardous
(classified) locations in accordance with drawing 2028219.
YCO16-Z Ex-Link Converter
Associated intrinsically safe apparatus for connection to Class I, Division 1 Groups C and D
hazardous (classified) locations in accordance with drawing 2003809 and Class I Zone 1 Group
IIB hazardous (classified) locations in accordance with drawing 2003809.
YCO17-Z Ex-Link Converter
Associated nonincendive field wiring equipment for connection to Class I, Division 2, Groups C
and D hazardous (classified) locations in accordance with drawing 2028219.
FM Approved for:
Sartorius Lab Instruments GmbH & Co KG
Gottingen, Germany
To verify the availability of the Approved product, please refer to www.approvalguide.com
FM Approvals HLC 5/13
116
Installation Instructions VIS1X
3055566
Page 2 of 3