Reviewing Test Results - Fora Diamond GD50 Manual Del Usuario

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 102
test strip to touch the drop. Once the confirmation window
is filled completely, the device will begin counting down.
Note:
To avoid contaminating the control solution, do not directly
apply the control solution onto a strip.
4. Read and compare the result. After counting down to 0, the
test result of the control solution will appear on the display.
Compare this result with the range printed on the test
strip vial or individual foil pack and it should fall within this
range. If the test result is out of range, read the instructions
again and repeat the control solution test.
(118 mg/dL = 6.5 mmol/L;
118-160 mg/dL = 6.5-8.8 mmol/L)
118~160
Note:
• Do NOT test your blood.
• The control solution range printed on the test strip vial or
individual foil pack is for control solution use only. It is not
a recommended range for your blood glucose level.
• Refer to the Maintenance section for important
information about your control solutions.
Out-of-range results:
If you continue to get results that fall outside the range
printed on the test strip vial, it means that the meter and
strips might not be working properly. Please contact your
local customer service or place of purchase for assistance.

Reviewing Test Results

Your device stores the 450 most recent test results along
with respective dates and times in its memory. To enter the
device memory, start with the device switched off.
To review all test results, do the following:
1. Press and release MAIN. The "
screen.
2. Press MAIN to review the test results stored in the device.
Press MAIN repeatedly to review other test results stored
in the device. After the last test result, press MAIN again
and the device will be turned off.
EN-10
" icon appears on the
Tabla de contenido

Solución de problemas

loading

Tabla de contenido