The message "ZERO SENSOR" is then displayed on the Pressio® monitor (PSO-3000).
The appearance of error codes, "E001", "E002", "E005" or the message, "CONNECT
SENSOR" on the Pressio® monitor after connecting the catheter to the extension cable
may be a sign of an incorrect connection.
In this case, make sure that the catheter connector is pushed right up to the stop
in the catheter extension cable, and that the latter is correctly connected
to the Pressio® monitor.
3) Preparation of the sterile water cup
Prepare a sterile saline solution (or even sterile water) and a cup in a sterile environment.
Pour the sterile saline solution (or sterile water) into the cup.
P
:
recautioN
p
RessuRe sensoR zeRoing must be peRfoRmed in shallow liquid
Receptacle could lead to hydRostatic pRessuRe gReateR than atmospheRic zeRo
would then lead to an eRRoneous RefeRence zeRo
4) Zeroing the catheter
Immerse the sensor (metal part) in the liquid completely.
Be careful not to touch the cup with the sensor.
When the sensor is ready, press the zeroing key.
(~5
). u
mm
sing a deep
,
which
.
N
:
ote
d
uRing zeRoing keep the sensoR immeRsed and avoid moving it in the cup
P
:
recautioNs
a
void making any contact with the end of the catheteR duRing sensoR zeRoing
c
alibRation in Relation to atmospheRic pRessuRe could be falsified
e
nsuRe the end of the catheteR is immeRsed in the steRile wateR oR a steRile saline solution
. s
duRing zeRoing
ensoR zeRoing in conditions otheR than those Recommended may cause
.
inaccuRate pRessuRe Readings
Movement of a task progress bar makes it possible to follow the correct performance
of the zeroing.
The procedure takes about 3 seconds.
5) Verification that the zeroing procedure took place correctly
When the progress bar reaches the end, the message "ZERO SENSOR OK – INSERT
SENSOR" shows that the procedure has taken place correctly.
Check that the Pressio® monitor screen displays the value "0 mmHg" (or the patient
monitor screen if the Pressio® interface is used).
If during zeroing, the sensor suffers parasite pressure variations or is damaged,
the screen will display the message "ZERO SENSOR FAILURE – REPLACE SENSOR".
In this case, disconnect and then reconnect the catheter, and try the zeroing sequence
again.
If this fails again, change the catheter.
Also refer to the instructions for use for the Pressio® monitoring system used
(PSO-3000 or PSO-IN00, § "Anomaly in operation: symptoms / suggested measures").
N
:
ote
s
ensoR zeRoing can only be done once
, "sensoR alReady zeRoed"
the message
Storing data in the catheter connector memory (cf. §3) makes it possible to use
Pressio® monitors or interfaces (PSO-3000 models software V2, PSO-3000 software
V1 and PSO-IN00) other than the model on which the zeroing was performed.
Specific case for the PSO-PBT:
N
:
ote
w
hen the tempeRatuRe Read by the sensoR is loweR than
: " - - - ". t
displays
his means that the measuRement is in the sensoR measuRement Range
but outside the display Ranges foR the
Implantation Technique
Positioning of the bolt and implantation of the Pressio® catheter must take account
of current aseptic neurosurgical practices.
Use the insertion accessories supplied by Sophysa in the Pressio® monitoring kit.