Wipe the LED Smart Stick thoroughly with a
ready-to-use, fast-killing disinfectant for the
specified contact time per manufacturer' s
instructions. Dry off the LED Smart Stick and
gold connector pins before use.
Surface Disinfection of the Power Hose
Zyris, Inc. recommends performing a
surface disinfection of the Power Hose
between each patient.
Use a hospital grade disinfectant per
manufacturer' s instructions. If components
are visibly soiled with biomatter or blood,
use an intermediate level disinfectant wipe
with a claim to inactivate Mycobacterium
tuberculosis. Barrier Sleeves may also be
used and changed between patients.
Cleaning and Sterilization of the
Vacuum Light Pipe
Zyris, Inc. recommends sterilization of the
Vacuum Light Pipe between each patient.
Remove biomatter and other debris from the
suction channels using the provided Double
End Tube Brush with a neutral pH (6.5-7.5)
enzymatic cleaning solution. Do not use
solution to hold parts. After cleaning, rinse
with tap water, dry, and place components in
an autoclave pouch to be sterilized in a
steam autoclave. Use validated sterilization
cycles (Table 4).
Cleaning and Sterilization of the
Titanium Control Base
Zyris, Inc. recommends following normal
infection control protocol for dental devices.
Users may sterilize the Titanium Control Base
at their discretion.
Remove the Vacuum Levers from the
Titanium Control Base. Remove biomatter
and other debris from the suction channels
using the provided Double End Tube Brush
with a neutral pH (6.5-7.5) enzymatic
cleaning solution. Do not hold parts in
cleaning solution or use ultrasonic cleaners.
After cleaning, rinse with tap water, dry, and
place components in an autoclave pouch to
be sterilized in a steam autoclave. Use
validated sterilization cycles (Table 4).
After sterilization, lubricate the Vacuum Lever
O-rings with the Dow 111 O-ring lubricant
provided in the O-ring Maintenance Kit.
Validated Sterilization Cycles
Cycle
Temperature
Standard
121 °C / 250 °F 30 min
Standard
132 °C / 270 °F 15 min
Prevacuum 132 °C / 270 °F 4 min
Prevacuum 134 °C / 273 °F 3 min
Table 4. Validated Sterilization Cycles
Min. Exposure Min. Drying
Time
Time
30 min
15 min
20 min
20 min