Disassembly of the Control Head
Step 1
Press the release button (Figure 7) on
the underside of the Titanium Control
Base to release and disconnect the
Vacuum Light Pipe.
Step 2
Remove the LED Smart Stick from the
Titanium Control Base.
Step 3
Holding the head of the Power Hose, pull
to remove the Titanium Control Base. Do
not twist.
Vacuum
Light Pipe
Vacuum Tubes
Release
Button
Vacuum
Levers
Vacuum Tubes
10
Led Smart Stick
(Do Not Autoclave)
Titanium
Figure 7. Control Head
Disinfection, Cleaning, and
Sterilization
Surface Disinfection of the Titanium
Control Base
Zyris, Inc. recommends performing a
surface disinfection of the Titanium Control
Base between each patient. To disinfect the
Titanium Control Base, first remove the
Vacuum Light Pipe and the LED Smart Stick.
If the Titanium Control Base is visibly
soiled with biomatter or blood, use an
intermediate level disinfectant wipe
with a claim to inactivate Mycobacterium
tuberculosis. Otherwise, use a hospital grade
disinfectant and follow the manufacturer's
instructions for contact time.
Surface Disinfection of the LED
LED Light
Control
Smart Stick
Button
Zyris, Inc. recommends performing a
surface disinfection of the LED Smart Stick
between each patient.
Spray a ready-to-use, fast-killing disinfectant
onto entire surface of the LED Smart Stick
and allow to remain wet per manufacturer' s
instructions. Using a nylon bristle brush,
Control
clean around the light control button area.
Base
Do not press the button while cleaning.
Gently work the bristles in a circular motion,
following the groove between the housing
and the button for approximately
60 seconds to remove all visual debris.
Repeat if any contaminants are still visible.