For servicing, only use the spare parts specified. The
manufacturer cannot be held liable for any accident caused
following a repair not undertaken by its own after-sales
service department or by an authorised repair technician.
12-1 Replacement of battery and fuses
The multimeter must be disconnected from any electrical
source.
Put the switch in the OFF position
Unscrew the captive screws and remove the lower half of the
casing
Replace the worn battery with a 9 V battery (6F22 or 6LF22)
Replace the faulty fuses, ensuring value and type are respected:
- 0.5 A HPC rapid F fuse (600 V – 10 kA – min 6.3 x 32 mm)
12-2 Storage
- 12 A HPC gR fuse (600 V min. – 300 kA – 10.3 x 38 mm)
If the multimeter is not to be used for a period of 60 days or more,
remove the battery and store separately.
12-3 Cleaning
The multimeter must be disconnected from any electrical
source.
To clean the case, use a cloth dampened with soapy water. Rinse with
a damp cloth. Then dry quickly with a cloth or pulsated air.
12-4 Metrological verification
As for all measurement instruments, periodic verification is
necessary.
For this instrument, we recommend at least an annual verification. For
the verification and calibration of your instruments, contact our COFRAC
accredited metrology laboratories or MANUMESURE branches.
Information and details on request:
Tel. : 02 31 64 51 43
12-5 Repairs under or outside of the warranty
Send your instruments to one of the CHAUVIN ARNOUX authorised
regional branches of MANUMESURE.
Information and details on request:
Tel. : 02 31 64 51 43
12-6 Repair outside mainland France
For any repair under or outside of the warranty, return the instrument
to your distributor.