Hot-air sterilization at 180°C
Hot-air sterilization of the inner chamber
Normal Display
3 sec.
GB
GB
Before starting the sterilization,
pull out the
CO
sensor without
2
any rotation from the connection
socket and remove it from the
inner chamber.
Empty the water
pan
.
The hot-air sterilisation takes10 to
12 hours in 3 steps:
- Rapid heating-up
- Maintaining the inner
temperature for 4 hours,
which
ensures 180°C
on
all internal
surfaces for at least 30 minutes
- Cooling-down phase up to 37°C
Start the sterilization cycle
Running sterilization
Alternating display
showing the actual temperature
Grenzert
Grenzert
Manual abort
of sterilization
Copyright
©
BINDER GmbH 10/2012
After the end of sterilization
10 to 12 hours
3 sec.
Ineffective sterilization when aborting prematurely!
Art. No. 7001-0235
Alternating display
Turn off the unit,
or press "180°C",
or open the outer door.
The unit is ready to operate.
Do NOT open the doors until the
interior temperature has
dropped to 37°C. After a manual
abortion, the C 150 reverts to its
standard operational status. As
long as the interior temperature
remains above the temperature,
which was set to trigger the
temperature alarm, the C 150
will trigger alarm signals. You
can switch them off by pressing
"EXIT". If necessary, repeat the
sterilization.