5. Slide the fan from the holding screws.
6. Clean the filter with soap and water and reposition.
7. Press ENTER.
8. Replace the fan. Make sure that the fan opening is positioned downward. Attach the
9. Close the analyzer enclosure and the analysis panel.
10. Press ENTER.
The instrument will reset the maintenance counter and will start the analyzer again.
6.1.3 Fuse replacement
The fuses for the power supply are found in the sc1500 controller. Refer to the sc1500
User Manual for fuse replacement information.
6.2 Reagent replacement
The chemicals must be changed or renewed at regular intervals. Refer to
information on the life of the chemicals.
Chemical
(section 8.1
on page
65)
Reagent
Standards (2 liters):
Cleaning and calibration
Electrolyte and
membrane cap
English
retaining strap (hold down the fan) and install the fan locking screw.
Table 2Chemicals for the AMTAX sc
Measuring range 1
Measuring range 2
(0.02–5 mg/L)
2500 mL for 3 month
2500 mL for 3 months
0.5 and 2.5 mg/L for
1 and 10 mg/L for
2 month on daily
3 months on daily
calibration
calibration
250 mL for 1 month on 3 cleanings and calibrations per day (default)
250 mL for 3 months on daily cleaning and calibration
11 mL
11 mL
Electrolyte for
Replace with
1–1.5 months;
membrane every
Membrane cap:
2–3 months
2–3 months
(depending on the
(depending on the
membrane soiling)
membrane soiling)
Measuring range 3
(0.05–20 mg/L)
(1–100 mg/L)
2500 mL for 2 months
10 and 50 mg/L for
3 months on daily
calibration
11 mL
Replace with
membrane every
2–3 months
(depending on the
membrane soiling)
Maintenance
Table 2
for
Measuring range 4
(10–1000 mg/L)
2500 mL for 2 months
50 and 500 mg/L for
3 months on daily
calibration
11 mL
Replace with
membrane every
2–3 months
(depending on the
membrane soiling)
15