5 OPERATION
5.1 Local Software Structure
5.2 Basic Operation
1) Select Test Mode
Press
2) Set Wavelength
Turn the Wavelength Setting Knob to select the wavelength you want, and the wa-
velength value can be displayed on the screen in real-time.
3) Input the Coefficient
When the system prompts you to input C, K or B, press
button to change the value till it displays the one you want, then press
to confirm.
Note: The system will always memorize your last input, until you input another value.
4) Calibrate Zero
Close the lid of the compartment and press
5) Calibrate 100%T
Pull the Reference in the light path, press
6) Print the result
Press the
Autovía A2 Km 585.1 Abrera 08630 (Barcelona) España
(Sujetas a modificaciones sin previo aviso) Pag.: 16
to select the measure mode.
button to print the test result.
or
to calibrate Zero T%.
to calibrate 100%T0.