Step 4
Read and Compare the Result.
After counting down to 0, the test result
of control solution will appear on the
display. Compare this result with the
range printed on the test strip vial and
it should fall within this range. If not,
please read instructions again and
repeat the control solution test.
Out-of-range results.
If you continue to have test results fall outside the range printed on
the test strip vial, the meter and strips may not be working properly.
Do NOT test your blood. Contact the local customer service or place of
purchase for help.
• The control solution range printed on the test strip vial is for control
solution use only. It is not a recommended range for your blood glucose
level.
• See the MAINTENANCE section for important information about your
control solutions.
NOTE
109~165