5.4
Optional programming
You can disable the start current check that otherwise checks that the resistance is not
higher than a specified limit. This disabling function is set in the MORE mode instead of
the SET mode in order not to confuse users who do not need this feature, and for safety
reasons.
The MORE mode is described in Fig. 5. After pressing MORE, the program asks you for
the program number and if you want to keep the start current check. Choose program
number and change between "YES" and "NO" by / and confirm with ENTER. Choosing
"NO" for start current check means that the function is disabled and high voltage can be
applied despite a very high resistance. Leave the MORE mode byEXIT.
WARNING! By disabling the start current check, the power supply can deliver high
voltage even if it is not connected to electrophoresis equipment.
1. Enter MORE
2. Select program
3. Choose current check no
4. Back to SET mode
Fig. 5. Disabling the current check.