General Description; Safety Precautions - GE EPS 601 Manual Del Usuario

2 General Description

Electrophoresis Power Supply EPS 601 User Manual 18-1130-19 Edition AD
2.1

Safety precautions

Extreme caution should be exercised in the operation of this instrument as it can
develop sufficient voltage and current to produce a lethal shock.
To avoid any risk of injury, the instrument should only be operated by properly trained
personnel and always in accordance with the instructions provided.
Read this entire manual before using this power supply.
1.
This instrument is designed for indoor use only.
2.
The instrument must always be used with the protective earth lead of the power
cord correctly grounded to earth at the mains outlet.
3.
To permit sufficient cooling, ensure that the vents in the rear and sides of the
instrument are not covered.
4.
Do not operate the instrument in extreme humidity (above 95%). Avoid
condensation by letting the unit equilibrate to ambient temperature when taking
the power supply from a colder to a warmer environment.
5.
Keep the instrument as dry and clean as possible. Wipe regularly with a soft damp
cloth. Let the power supply dry completely before use. If wetted, unplug the power
supply until the instrument is dry.
6.
Use only undamaged electrical wire and equipment specified for the voltages you
will use. High voltages wires must meet the requirements of the
IEC 1010-2-031:1993 electrical standard. Any electrophoresis equipment connected
to the power supply should meet the requirements of the IEC 1010-1:1993
electrical standard.
7.
Note that the output is connected to the chassis/reference earth.
General Description 2
9
loading