en
Position connector for control unit c on the motor cable so that mark-
ing b on the connector is aligned with marking a on the connection
socket for applied parts 7, see Fig. 3.
Plug connector for control unit c on the motor cable firmly into one of
the two connection sockets for applied parts 7.
Fig. 3
Connecting the motor cable
Legend
a
Marking for connection socket
b
Marking for connector
c
Connector for control unit
Note
The motor cable must be connected to the control unit without an applied
part or with the applied part locked ("Off" position).
Otherwise the control unit will not recognize the applied part, and a mes-
sage will appear on the display (except for drill tool GA844).
If the applied part is not recognized;
– Lock the applied part, see Chapter 6.1.12.
– Release the applied part again, see Chapter 6.1.11.
6.1.9

Connecting the ELAN 4 electro single-use tube set GA395SU

Note
The tube set is sterile. Sterile separation occurs at the tube set.
Open flap a of the coolant and irrigation pump 3, see Fig. 4.
Insert single-use tube set c:
– Place pump tube with loop over the roller wheel b.
– Push nib e of the single-use tube set under the pull slide d until the
nib engages.
Close flap of the coolant and irrigation pump a. When doing this, make
certain the pump tube does not get clamped in.
Insert the bottle holder 16 for the sterile fluid bottle in bottle holder
mount 15.
Insert the punching spike into the sterile fluid bottle.
When using sterile fluid bottles made from glass: Open the ventilation
flap on the punching spike.
Hang the sterile fluid bottle from the bottle holder 16.
Fasten the tube to the motor cable using fixation clamps.
Shorten the length of the tubes to correspond with the applied part
being used and connect to the spray nozzle.
10
a
b
c
Fig. 4
Connecting the single-use tube set
Legend
a
Flap of the coolant and irrigation pump
b
Roller wheel
c
Single-use tube set
d
Pull slide
e
Nib of the single-use tube set

6.1.10 Connecting the applied part to the motor cable

Note
All ELAN 4 electro motor cables (GA804/GA805/GA806) can be used in
combination with all applied parts that do not have a fixed connection with
a cable.
Connect connector for motor cable 24 to connector for application
part 19 of the motor cable. Ensure that the nose 23 on the motor cable
is aligned with the groove on the coupling of the application part.
The applied part engages. When the visual field is "Off" 20 on the
motor cable, a gold-colored marking is visible.
The control unit 1 recognizes the applied part type and displays this
type in the corresponding applied part control panel in the display 2.
The most recent settings for this applied part type at the same motor
connection socket are shown on the display 2.
Note
The applied part connected to this motor cable is only ready for operation if
a gold-colored marking is visible in the visual field "On" 21 on the motor
cable.