1
2
30290627D PFA-series
5. Pull the load plate up using the handle.
DANGER
Danger of injury due to slamming load plate.
▲ Ensure that the pneumatic spring has extended completely.
Securing the raisable load plate
1. On the right hand side of the weighing platform slide the slot of the supplied safety
wedge (1) onto the load plate.
2. Slide the safety wedge down until it will go no further.
3. Make sure that the catch (2) is on the inner side of the load frame.
The raised load plate is secured and cleaning or service work can be performed safely.
13
Maintenance