Cleaning and Disinfection
CAUTION
• Do not immerse the nebulizer unit in any liquid and do
not hold it under running water. Be careful to ensure that
no water penetrates the nebulizer unit.
• Do not clean the nebulizer or its component parts in a
dishwasher.
• Do not allow any moisture to come into the battery com-
partment.
• Do not use any other disinfectant like described above to
clean the unit or its components.
• Do not use a hairdryer to dry any parts.
• Do not clean the unit and the accessories with brine
(NaCL).
Maintenance
When using a new mesh cap for the fi rst time, and after
changing the batteries you need to reset the vibration rate:
GB
1. Pour water into the medicine cup (pay attention to maxi-
mum marking).
2. Attach the mesh cap to the nebulizer unit.
3. Take the batteries out.
4. While holding down the ON/OFF button, reinsert the bat-
teries and close the battery cover.
5. Keep the unit upright and then release the ON/OFF but-
ton. The green and red indicator light up, indicating that
the nebulizer is resetting the vibration rate.
6. After a few seconds, the nebulizer automatically turns off.
The vibration rate resetting is fi nished. The nebulizer is
ready for use.
The unit must be kept upright during this procedure.