Cleaning and Disinfection
Daily disinfecting
Disinfection by boiling:
1. Silicon masks and mesh cap can be disinfected by boil-
ing in distilled water.
2. After 10 - 20 minutes of boiling, gently wipe off excess
water with a clean dry cloth and allow to air dry on a
clean paper towel.
Disinfection by using alcohol:
1. PVC mask, mouthpiece, unit lid and mesh adapter can
be disinfected by using medical alcohol (Ethanol 75%
dilution with water).
2. Disinfect the above-named parts by wiping with a cloth
moistened with medical alcohol (Ethanol 75% dilution
with water).
3. Let them air dry on a clean paper towel.
Note: Only silicon mask can be disinfected by boiling,
PVC mask just can be disinfected by using medical alco-
hol 75 % (dilution with water).
Cleaning and Disinfection the Nebulizer Unit
Cleaning after every use
1. Use a cloth moistened with water to clean the medication
cup and outer casing of the unit.
2. Wipe off the unit with an unused dry cloth and let it air
dry.
Daily disinfecting
1. Use a cloth moistened with medical alcohol (Ethanol
75 % dilution with water) to disinfect the medication cup
and outer casing of the unit.
2. Let them air dry on a clean paper towel.