Please read all the information in this manual before using the BL
WARNING
Contact with low-temperature objects may cause cold burns. Users should be aware of the following
potential hazards:
• USE CARE AND WEAR PROTECTIVE GLOVES TO PROTECT HANDS
• DO NOT place any liquid directly into the instrument.
OPERATOR SAFETY
All operators of Techne equipment must have available the relevant literature needed to ensure their
safety. It is important that only suitably trained personnel operate this equipment, in accordance with the
instructions contained in this manual and with general safety standards and procedures. If the equipment
is used in a manner not specified by Bibby Scientific Ltd. the protection provided by the equipment to the
operator may be impaired.
All Techne instruments are designed to conform to international safety requirements and are fi tted with
an over-temperature cut-out. If a safety problem should be encountered, switch off the unit at the mains
socket and remove the plug from the electricity supply.
INSTALLATION
The instrument should be carried using both hands. Never move or carry the instrument when in use or
connected to the mains electricity supply.
1. All Techne instruments are supplied with a power cable; this may be integral or plug-in.
2 .
Before connecting the instrument to the mains electricity supply, check the voltage
against the rating plate (located on the back of the unit). Please note that the unit
must be earthed to ensure proper electrical safety. Connect the mains cable to a
suitable plug according to the table below.
Connection
Live
Neutral
Earth
3. UK ONLY: The fused plug supplied with the mains cable is fitted with a 5 amp fuse to protect
the instrument and the operator.
4. The units are rated to operate at 100-230V, 50/60Hz.
5. Place the unit on a suitable flat bench or in a fume cupboard if required, ensuring that the air
vents on the rear and underside are free from obstruction.
SAFETY AND INSTALLATION
220/240V, 50/60Hz Supply
Brown
Blue
Green/yellow
4
o
CKICE.
110V/120V Supply, 50/60Hz
Black
White
Green