Programming - HACH LANGE si628 C Manual De Instrucciones

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

4. Operation

4.4. Programming

The instrument is supplied with the default configuration specified in Setting up, page 14.
To modify the default configuration and adapt it to specific applications follow the diagram below.
MEASURING 22.3°C
545 µS/cm
@ 25°
CODE 000
Enter code 100.
Select the nominal value of cell
constant to be used. If the exact
value is known, it can be directly
entered, avoiding the calibration
procedure.
Selection of the functioning
values of the corresponding
relays.
Input "ON" (closed contact).
Output "OFF" (open contact).
Selectable values between
0.0 µS and 200 mS.
Select the time before the alarm
activation.
Selectable between 1 to 99 min
or hours. 0= deactivated.
Select the conductivity values
corresponding to the Current
Output (4 mA and 20 mA).
Selectable values between
0.0 µS and 200 mS.
Selection of the current output
when the instrument is in Hold
state: fixed at 21 mA or at the
mA corresponding to the last
measured value.
Selection of the working
language: English, German,
Spanish, French, Italian,
Portuguese.
(*) LIMIT and OUTPUT 4-20 mA:
Allows the introduction of five
digits, placing the decimal point
you want.
16
si628 C_HACH LANGE_Ed. 0510.indd ANG:16
si628 C_HACH LANGE_Ed. 0510.indd ANG:16
Note: When programming, the instrument is
in hold state:
• The contact relays are open.
• The current output is frozen at 21 mA.
To quit the hold state it is necessary to return
to MEASURE.
CALIBRATE
CELL
Vary the underlined digit with
LIMIT (*)
Vary the underlined digit with
ALARM
OUTPUT 4-20 mA
Vary with
HOLD
Vary with
LANGUAGE > Español
LANGUAGE
MEASURING
MEASURE S/cm
CONST:
1.000 cm
–1
and accept with
ON - - - - - - OFF
1000 µS
800 µS
and accept with
ALARM:
01 min
4 mA
20 mA
pH: 0.0
2.0 mS
> 21 mA
last
Italiano
22.3°C
545 µS/cm
@ 25°
20/5/10 11:08:45
20/5/10 11:08:45
Tabla de contenido
loading

Tabla de contenido