Cell Cables; Cells With Mp-5 Connector - HACH LANGE si628 C Manual De Instrucciones

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

3. Installation
Terminal assignment (rear panel)
CRISON INSTRUMENTS, S.A.
E-08328 ALELLA - Barcelona
Made in Spain
4-20 mA
RED
VCA

3.3. Cell cables

Compensation of the cable length
To minimize the effect of the cable resistance in a conductivity measurement, the cable length is compensa-
ted by the 4-wire system. With the Pt 1000 temperature sensor, the cable resistance has little importance.
However, in order to reduce the resistance of the conductor to a half, 2 cables have been connected to
the ends of the sensor.

3.3.1. Cells with MP-5 connector

MP-5 connector, 5 contacts
12
si628 C_HACH LANGE_Ed. 0510.indd ANG:12
si628 C_HACH LANGE_Ed. 0510.indd ANG:12
Transmitter si628 C
S/N:
IP 54
CELL
L 1
L 2
Notas importantes:
• El si628 C does not require ground connec-
tion (terminal 1).
• Connect the ground only when there could
be interference problems.
DANGER
Relays:
- Limit the current up to 3 A by using
one fuse.
- Do not connect cables with voltage
higher than 24 V.
TERMINALS
si628 C
si628 C
(wall mount)
(panel mount)
6
6
7
7
8
9
10
11
12
COLOUR
16
White
16
Brown
15
Pink
15
Yellow
14
Shield
13
Green
12
Blue
11
Gray
10
Red
20/5/10 11:08:43
20/5/10 11:08:43
Tabla de contenido
loading

Tabla de contenido