7.

CLEANING

The headlamp module can be wiped down with alcohol. The headlamp module
lens and the camera window should be cleaned only with lens tissue, available in
any camera store, by following the directions on the package.
Follow all applicable bloodborne pathogen procedures as required by OSHA and
or your hospital, when cleaning, disinfecting and sterilizing.
The sweatband may be cleaned using a standard wash cycle in warm water and
detergent.
The headband can be cleaned with an alcohol wipe down or washed with a
luke warm solution of water and mild detergent. Rinse thoroughly.
Sterilizing of lensed fiber optic cables should be avoided, the lens at the end of
the cable is not sealed and moisture will enter between the lens and the cable.
WARNING: Do not autoclave the headband, lensed cables or sweatband.

8. REPLACEMENT PARTS

Item
Fiber Optic cable, Bifurcated with lens ................
FiberOptic cable, Single with lens........................
Carrying Case ...............................................
Garment Clips (Quantity 5) ..............................
Comfort Liner, Set ...........................................
LIT-092 CUDA
ELITE
®
(English)
Part number
SYC3500B-BLUE-10
SYC3500S-BLUE-10
SYC0063
SYC0062
SYC-0040
Rev. E