This instrument has been designed to meet the requirements of the EMC Directive 89/336/EEC.
Compliance was demonstrated by meeting the test limits of the following standards:
Emissions
EN61326 (1998) EMC product standard for Electrical Equipment for Measurement, Control and
Laboratory Use. Test limits used were:
a)
Radiated : Class A
b)
Conducted : Class A
c)
Harmonics: EN61000-3-2 Class A; the instrument is Class A by product category
according to Amendment 14.
Immunity
EN61326 (1998) EMC product standard for Electrical Equipment for Measurement, Control and
Laboratory Use.
Test methods, limits and performance achieved were:
a)
EN61000-4-2 (1995) Electrostatic Discharge : 4kV air, 4kV contact, Performance A.
b)
EN61000-4-3 (1997) Electromagnetic Field, 3V/m, 80% AM at 1kHz, Performance A.
c)
EN61000-4-11 (1994) Voltage Interrupt, 1 cycle, 100%, Performance B*.
d)
EN61000-4-4 (1995) Fast Transient, 1kV peak (AC line), 0.5kV peak (RS232 port),
Performance A.
e)
EN61000-4-5 (1995) Surge, 0.5kV (line to line), 1kV (line to ground), Performance A.
f)
EN61000-4-6 (1996) Conducted RF, 3V, 80% AM at 1kHz, AC line, Performance A.
* In a controlled EM environment as defined by EN61326 Annexe B (½ cycle, 100% interrupt) the
TG420 will meet Performance A criteria.
According to EN61326 the definitions of performance criteria are:
Performance criterion A: 'During test normal performance within the specification limits.'
Performance criterion B: 'During test, temporary degradation, or loss of function or
performance which is self-recovering'.
Performance criterion C: 'During test, temporary degradation, or loss of function or
performance which requires operator intervention or system reset occurs.'
Cautions
To ensure continued compliance with the EMC directive the following precautions should be
observed:
a) connect the generator to other equipment using only high quality, double-screened cables.
b) after opening the case for any reason ensure that all signal and ground connections are
remade correctly before replacing the cover. Always ensure all case screws are correctly
refitted and tightened.
c) In the event of part replacement becoming necessary, only use components of an identical
type, see the Service Manual.