The adjustable dynamics parameters can be selected (Fig. 153).
Fig. 153
4.6.5.3 Attenuation setting
The pH or mV signal becomes essentially quieter after a specific setting period when the attenuation is switched
on («low», «medium» or «strong») (Fig. 154).
Fig. 154
A minimum waiting period should therefore also be observed for the various attenuation settings.
Attenuation setting
None
Weak
Average
Strong
Minimum waiting period
1 second
2 - 3 second
3 - 4 second
5 second or more
Application
All aqueous titration applications
Titrations in polar solvents such as ethanol
Titrations in partially nonpolar solvents of ethanol/toluene
Titration in non-polar solvents or harsh applications such
as TAN