Xylem SI Analytics TitroLine 7800 Manual De Instrucciones página 217

Titulador
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

217
Fig. 148
If normal or fast drift was selected before, the values will be defaulted for user-defined drift. In the present case,
for instance, 20 mV for normal drift: (Fig. 149).
Fig. 149
Drift-controlled acceptance of the measured value is used in most applications.
However, there are applications in which the setting of a fixed holding time for measured value acceptance
following the titration step is recommendable. Examples hereof include titrations in non-aqueous media. In the
case of Dead-Stop titration no holding time other than the fixed one can be selected. The fixed delay time can be
set between 0 and 999 seconds (Fig. 150).
Fig. 150