• General installation and starting tips
• Installation
γ
Aspi-Jet 7
is fitted with automatic drainage, it must be connected to the waste of the building.
Besides the general features of Aspi-Jet 7
function: water supply to the tumbler and spittoon (cuspidor) with rinsing respectively. The
switches which control the above functions are marked with symbols and are located on the front
panel:
- a tumbler indicates the water supply on type 8
- a tap indicates rinsing of the spittoon on type 9
• General installation and starting tips
• Unpack the appliance following the instructions shown on the package.
• Dispose of the package in compliance with regulations.
• Verify that the appliance has not been damaged during transport.
• Do not connect damaged appliances to the mains.
• Do not use extension leads, multiple plugs or sockets.
• Ascertain that the feeding line is adequate to feed the machine.
• Assembly of the aspirator must be carried out by an expert, with suitable equipment and special
training. The installer should read the equipment manual, perform commissioning and instruct
the users in the use and routine maintenance of the new machine while it is still unused and
therefore not contaminated.
• The air exhausted by the aspiration system should be filtered, with a special certified antibacterial
filter available on request, and expelled to the outside.
• After installation, perform the required safety and operating tests.
• Arrange for periodical inspection of the equipment. This will not only prevent stoppage of a chair
or of the whole surgery, but is also a way to prevent injuries and accidents.
• You can find all our updated manuals at our website: www.cattani.it. We recommend that you
consult them, especially for updates on the subject of safety.
• Installation
The aspirator must be installed in compliance with CEI 62-5
regulation for electro-medical appliances (Aspi-Jet 6-7-
γ
8-9
have been designed accordingly).
Prior to plugging in the unit, check the specifications on the
label and make sure that the mains are compatible with the
appliance and protected against overcurrent according to
CEI 64/8 regulations.
The appliance must be protected against indirect contacts
for class
appliances according to CEI 64-8 reg. and 64-4
for rooms used as medical consulting rooms (IEC cor-
respondents available on demands).
The plug and cable are equipped with earthing protection:
do not remove this protection in any case and make sure
that the socket is compatible.
18
γ,
γ
Aspi-Jet 8
and Aspi-Jet 9
γ
;
γ
.
γ
offer some additional
3C
2