Place the furnace on a horizontal surface, flat and firm, away from vibration sources. Make sure there is
▪
enough room for ventilation.
Make sure the casters of the furnace are braked or locked during operation.
▪
The closest element should be at least 15 cm away from the unit.
▪
Connect the unit to a grounded 230 V, 50/60 Hz power source. It is also important that the power source is
▪
protected by a thermal magnetic circuit breaker of an appropriate voltage (consult the installation
engineer).
Before operating the furnace, it should be pre-heated in order to evaporate the humidity of the muffle. To
▪
do this, program the furnace to keep a temperature of 300 ºC for an hour.
BASIC PROGRAMMING
The electronic controls of the R-080110 allow to program up to 50 steps. Each step consists of two values:
C
Temperature at which the furnace will start each step.
t
Time selected to reach the next temperature.
The parameters for the example graphic (Fig. 9) would be the following:
Step
Parameter
C1
Step 1
t1
C2
Step 2
t2
C3
Step 3
t3
C4
Step 4
t4
C5
Step 5
t5
C6
Step 6
t6
12
Value
Meaning
0 ºC
Initial temperature
30 min.
Time selected to reach next temperature
400 ºC
First set-up temperature
15 min.
Time selected to reach next temperature (maintenance)
400 ºC
First set-up temperature (maintenance)
30 min.
Time selected to reach next temperature
1000 ºC
Second set-up temperature
15 min.
Time selected to reach next temperature (maintenance)
1000 ºC
Second set-up temperature (maintenance)
15 min.
Time selected to reach next temperature
600 ºC
Third set-up temperature
-121
End
Fig 9