Basic Programming - Mestra R-080110 Manual Del Usuario

Horno de circonio
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

English
Place the furnace on a horizontal surface, flat and firm, away from vibration sources. Make sure there is
enough room for ventilation.
Make sure the casters of the furnace are braked or locked during operation.
The closest element should be at least 15 cm away from the unit.
Connect the unit to a grounded 230 V, 50/60 Hz power source. It is also important that the power source is
protected by a thermal magnetic circuit breaker of an appropriate voltage (consult the installation
engineer).
Before operating the furnace, it should be pre-heated in order to evaporate the humidity of the muffle. To
do this, program the furnace to keep a temperature of 300 ºC for an hour.

BASIC PROGRAMMING

The electronic controls of the R-080110 allow to program up to 50 steps. Each step consists of two values:
C
Temperature at which the furnace will start each step.
t
Time selected to reach the next temperature.
The parameters for the example graphic (Fig. 9) would be the following:
Step
Parameter
C1
Step 1
t1
C2
Step 2
t2
C3
Step 3
t3
C4
Step 4
t4
C5
Step 5
t5
C6
Step 6
t6
12
Value
Meaning
0 ºC
Initial temperature
30 min.
Time selected to reach next temperature
400 ºC
First set-up temperature
15 min.
Time selected to reach next temperature (maintenance)
400 ºC
First set-up temperature (maintenance)
30 min.
Time selected to reach next temperature
1000 ºC
Second set-up temperature
15 min.
Time selected to reach next temperature (maintenance)
1000 ºC
Second set-up temperature (maintenance)
15 min.
Time selected to reach next temperature
600 ºC
Third set-up temperature
-121
End
Fig 9