Units
Converts the display from mils to microns and vice versa.
Reset
Reset (menu reset) restores factory settings and returns
the Gage to a known condition. The following occurs:
- All batches, stored measurements, batch names and screen
captures are erased.
- The Zero adjustment is returned to the factory setting.
Since this may not always be a precise zero, the Gage
should be checked on the glass plate.
- Menu settings are returned to the following:
Memory = OFF
Statistics = OFF
Hi Lo Alarm = OFF
Auto Dim = ON
Perform a more thorough Hard Reset as follows:
1. Power down the Gage and wait 5 seconds.
2. Simultaneously press and hold the ( + ) and
buttons until the Reset symbol
This returns the Gage to a known, "out-of-the-box" condition.
It performs the same function as a menu Reset with the addition of:
- Bluetooth Pairing info is cleared.
- Menu settings are returned to the following:
Units = microns
Flip Display = Normal
Sound = ON
NOTE:
- Keep the gage away from metal during a Reset.
- Date, Time and WiFi settings are not affected by either Reset.
Battery Type
Selects the type of batteries used in the Gage from a choice of "Alkaline" ,
"Lithium"or "NiMH"(nickel-metal hydride rechargeable). If "NiMH" is selected,
"Lithium"or "NiMH"(nickel-metal hydride rechargeable). If "NiMH" is selected,
the Gage will trickle charge the batteries while connected via USB to a PC
or optional AC power supply (USBAC). The battery state indicator symbol
is calibrated for the selected battery type. No damage will occur if the
battery type used in the Gage does not match the selected battery type.
Setup Menu
Bluetooth & Stream = OFF
WiFi & Access Point = OFF
Display = None
Language = English
Battery Type = Alkaline
Backlight = Normal
Bluetooth Smart = OFF
USB Drive = ON
4
appears.
center