8. Replacing the chuck
8.1 Remove the motor cable 15. To do so unscrew
the cap 16 from the motor and loosen the
cable by pulling out the plug-in seal 17.
8.2 Open the chuck and withdraw the rotary
instrument. Unscrew the end of the handpiece
21 (right-handed thread).
8.3 Remove seeling cap 24
8.4 Use a number 6223 wrench 23 to hold the
motor end of the spindle. Engage the triangular
section of the chuck with a No. 4115 chuck
tool 22. To prevent the chuck becoming distorted,
an instrument shank or the pin supplied must
always be inserted into it. Hold the number 6223
23 wrench tightly and turn the chuck tool anti-
clockwise to unscrew the chuck completely -
the chuck has a right-handed thread!
6653/1
6652
6652/1 Ø 3,00 mm
6642 Ø 2,35 mm
6643 Ø 3,00 mm
7052
4115
7250
6223
9. Care and maintenance
Caution: Do not clean your C3 Master
motorhandpiece with compressed air!
Chuck: should be cleaned and re-greased once in
a while, depending on how dirty it is.
As C3 Master units have no commutators, carbon
brushes or ventilation parts, no further maintenance is
required.
Ø 2,35 mm
8.5 Clean the chuck, grease it lightly and place it in the
spindle. Use the tools to screw the chuck in - clock-
wise and as far as possible - as described in section
8.4, and tighten it slightly.
Replace the plug-in seal 17 and screw the cap 16
back into place. Screw the sealing cap 24 and the
end of the handpiece 21 on again.
Please note:
The tools 22 and 23 are to be found in the
base of the handpiece rest!
10. Possible faults
Should the handpiece be overloaded, respectively,
jammed, for safety reasons the unit will switch itself
off.
- Release knee/foot switch and operate it again; the
unit is ready for use immediately
bench controller version only
- When using the bench version without a pedal
switch, the speed control knob should be turned to
"0", when the desired speed is selected again; the
unit is ready for use immediately.
- When using the bench version with a pedal switch
it should be released and operated again; the unit is
then ready for use again.
all versions
if the unit should not be running after doing as
described above the 2 fuses have to be checked and
eventually exchanged.
The fuses (2x T2AH250V/2x T4AH100V-115V)are to
be found at the rear of the unit beneath the socket 1.