10. Insert the mercury bulb ④ (respect the polarity of
the bulb), tighten the locking screws and refit the
lamp holder inside the lamp housing. (Fig. 20)
•
Do not touch the bulb with bare hands.
11. Connect the cable from the external power sup-
ply to the lamp housing. (Fig. 21)
12. Connect the power cable to the external power
supply. (Fig. 22)
•
The input voltage for fluorescence
power supply is 110-240Vac.
•
Please use the standard power ca-
ble provided. Select suitable one
when missing or damaged.
•
Connect the power supply cor-
rectly, be sure to have a good earth
connection.
•
Before connecting the power cord,
secure the lamp housing cable to
the power supply.
•
If the power cord is connected be-
fore, ther may be a risk of electrical
shock.
•
Disconnect all electrical cables
before installing or replacing the
bulb.
•
The bulb has an anode and a
cathode of different sizes. Respect
the polarity during assembly.
Page 15
④
Fig. 20
Fig. 21
Fig. 22