OHAUS Frontier FC5513 Manual Del Usario página 31

Ocultar thumbs Ver también para Frontier FC5513:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

TM
Frontier
Centrifuge FC5513
12.4 Table 4: Error messages
Error-No.:
Description
1
Imbalance arose
2
Imbalance sensor is defective
4
Imbalance switch has been activated for longer than 5 seconds
8
Transponder in the rotor is defective
14
Leap of speed is too big between two mesaurements
CLOSE lid
15
Standstill monitoring defective
16
Wrong dircention of rotoration of the motor
33
Open lid while motor is running
34
Lid contact defective
38
Lid motor is blocked
40
Communication with frequency converter distrubed during start
41
Communication with frequency converter distrubed during stop
42
Short circuit in the frequency converter
43
Undervoltage frequency converter
44
Overvoltage frequency converter
46
Over temperature motor
47
Over current frequency converter
48
Timeout between control unit and frequency converter
49
Other error frequency converter
55
Overspeed
70
Timeout between controler and RS232 interface
80
Memory Error intern EEPROM
99
Rotor is not allowed in this centrifuge
Inserted rotor does not exist in the programm
FALSE
Rotor is not detected
rotor no
EN-25
Tabla de contenido
loading

Tabla de contenido