12.3. Status messages
Message
01 MANUAL START
02 REMOTE STOP
03 REMOTE START
12.4. Error messages
Message
-01 TEMP / LEVEL ALARM
-03 EXCESS TEMPERATURE
WARNING
-04 LOW TEMPERATURE WARNING
-05 TEMPERATURE MEASUREMENT
ALARM
-07 I 2 C-BUS WRITE ERROR
-07 I 2 C-BUS READ ERROR
-07 I 2 C-BUS READ/WRITE ERROR
-08 INVALID COMMAND
-10 VALUE TOO SMALL
-11 VALUE TOO LARGE
-12 WARNING : VALUE EXCEEDS
TEMPERATURE LIMITS
-13 COMMAND NOT ALLOWED IN
CURRENT OPERATING MODE
82
Description
Waterbath in keypad control mode.
Waterbath in „OFF" state
Waterbath in remote control mode
Description
Safety temperature or low liquid level alarm
High temperature warning „
Low temperature warning „
Error in measuring system
Internal error
Invalid command
Entered value too small
Entered value too large
Value lies outside the adjusted range for the high
and low temperature warning limits. But value is
stored.
Invalid command in current operating mode
„.
„.