First turn off the monitor. Keep the select button
pressed, until a tone sounds. The display shows a
flashing «OK» next to the position sensor symbol (see
illustration).
If you wish to turn off the position sensor, press
the «I/0» button (C): The display changes from
«OK» (on) to «–» (off). Press the memory button to
confirm the new setting. Then the monitor turns off
automatically. In order to re-activate the position
sensor, follow the same steps as described above.
Storage, cleaning and disinfection of
the product
Cleaning and disinfection of the device housing and
the cuff may be done with commercially available
cleaning and disinfection solutions. Wipe cuff and
housing with a swab that is slightly moistened with
the solution and immediately wipe dry with a clean
cotton swab.
Do not store the product in a place exposed to direct
sunlight or high humidity.
Do not submerge the unit in water or any other liquid.
Calibration
This device has been calibrated at the time of
manufacture. If used according to the use
instructions, periodic re-calibration is not required.
Manufacturing date is given by the LOT number on
the back of the product. The first number after LOT
represents the last digit of the year of manufacture.
The next three digits give the day of the year of
manufacture.
An example: LOT 0116 xx xx... this product has been
manufactured on the 116th day of the year 2000.