Cleaning The Transducer - Vyaire SpiroUSB Manual De Funcionamiento

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 132

Cleaning the Transducer

The transducer requires no routine maintenance or servicing.
However, if you wish to disinfect or clean the transducer it may be
removed by means of the following procedure:
1. Rotating the turbine transducer anti-clockwise until the
locating pip lines up with the small rectangular cut-out in
the housing as shown below.
2. Gently pull the transducer away from the housing.
3. The transducer may now be immersed in warm soapy water
for routine cleaning or immersed in a cold disinfectant
solution e.g. Perasafe (Cat No. 36-SSC5000A) for a
maximum of 10 minutes (Alcohol and chloride solutions
should be avoided).
4. After cleaning/disinfecting, the transducer should be rinsed
briefly in distilled water and dried.
5. Re-assemble the mouthpiece holder.
Turbine transducer
CAUTION: Do not attempt to wash or immerse the SpiroUSB
transducer housing in water or cleaning fluid, as there are electro-
nic components inside that will be permanently damaged
Housing
Locating Pip
Cut-Out
25
Tabla de contenido
loading

Tabla de contenido