FujiFilm SonoSite L52 Serie Guia Del Usuario página 12

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 62
Table 2: TI (TIS, TIC, TIB)
Transducer
L52n (NanoMaxx)
L52x (M-Turbo or MicroMaxx)
L52x (S Series)
L52e (MicroMaxx)
L52 (TITAN)
L52 (180PLUS)
↓ Decrease or lower setting of parameter to reduce TI.
↑ Raise or increase setting of parameter to reduce TI.
— Data are not applicable for this transducer/mode.
Transducer surface temperature rise
Table 3
lists the measured surface temperature rise (in °C) from ambient (23°C ± 3°C) of transducers
used on the ultrasound system. The temperatures were measured in accordance with EN 60601-2-37
section 42 where controls and settings were positioned to give maximum temperatures.s
Table 3: Transducer Surface Temperature Rise IEC 60601-2-37 (Internal Use)
Test
Still air
Simulated use
6
Color Power Doppler
Settings
PRF
7.5
8.8
5.6
5.9
PW Settings
Depth
Sample Vol ↓
Sample Vol ↑
8.2
13.0
5.6
5.5
9.3
10.8
2.4
2.4
Tabla de contenido
loading

Tabla de contenido