Phosphate -ortho- (LR) 0 - 4 mg/l
(44)
Continued...
10) Press button 4 to save the result together with date and time in the
4
device.
!
Assessment of the measurement value relative to the ideal range defined by you.
Measurement value for this parameter is outside the value range of this method.
Note(s)
->
(*) Conversion to P, P O
->
With this procedure ortho-phosphate ions are detected. Other phosphates
must therefore be converted to ortho-phosphates before the test is begun.
->
The pH value of the sample water should be between 6 and 7 pH.
->
The following contents of substances in the sample water can - at the
respective concentration - falsify the measurement results:
Chromium > 100 mg/l
Copper > 10 mg/l
Iron > 100 mg/l
Nickel > 300 mg/l
Zinc > 80 mg/l
202
Name on device: 44-Phosphat-LR-tab
Possible messages in the line below the measurement value:
"low! / high! / good!":
"Overrange! / underrange!":
2
5
(PO )
Tablet
4
202