Iron in Oil
(149)
Adapter
Syringe
Measurement procedure:
The steps up to the selection of the parameter value to be determined are the same for all
procedures and described on page "TEST-5". Please observe the important notes for
accurate measurements on pages "TEST-8" and "TEST-9".
!
The individual steps of the measurement procedure are shown on the display.
Fill a clean 3ml vial with 2ml of Reagent S.
1)
Place 3ml vial-adapter into PrimeLab.
2)
3) Place 3ml vial into adapter and aplpy light shield .
4) Press ZERO.
4
One-Time-Zero:
Measure new ZERO
Use last ZERO
1
2
3
!
5) Remove light protection shield and 3ml vial.
6)
Handle the sample according to the TriboIron pill procedure.
Close the lid and shake beaker vigorously for 02:00 minutes.
7)
8)
Press OK to start 02:00 minutes countdown.
Let beaker rest for 08:00 minutes.
9)
10)
Press OK to start 08:00 minutes countdown.
11) Fill a clean 3ml vial with 2ml of the
solution.
133
20 - 550 mg/l
Name on device: 149-Iron-Oil-liq
(*)
(*)
Filter paper Filter adapter
Use button 3 (arrow down) to scroll through the steps.
Use button 4 to skip the notes.
If after switching on the device you have already performed the
ZERO (baseline) adjustment for the same type, the following
message is displayed:
-> Measure new ZERO
-> Use last ZERO
4
For the sake of precision select "Use last ZERO" only
if the stored ZERO value is used with the
same cuvette and the same water sample for the test!
2+
(Fe )
(*)
beaker
= continue with step 10)
= continue with step 11)
purple filtrated aqueous
Tablet
Liquid
+Liquid
TM-reagent-S
(TM149reagS)
TM-reagent-E
(TM149reagE)
TM-reagent-I
(TM149reagI)
Continued...
133