DEWAN-50
(109)
Continued...
!
Assessment of the measurement value relative to the ideal range defined by you.
Measurement value for this parameter is outside the value range of this method.
B) Range 150 - 300 mg/l DW-50
1) Fill 10ml of the sample water into a clean 24mm cuvette.
One-Time-Zero:
Measure new ZERO
Use last ZERO
1
2
3
!
2) Replace the screw lid on the cuvette and insert it into the PrimeLab
sampling chamber. Align the arrow on the front of the cuvette. Press
4
ZERO.
3) Remove the cuvette from the chamber and unscrew the lid.
4) Add 40 drops "PL-DX DEWAN-50" liquid reagent to the test liquid in
the cuvette.
5) Replace the lid on the cuvette, swivel it carefully and insert it in the
sampling chamber of the PrimeLab. Align the arrow on the front of the
4
cuvette. Press TEST.
6) After the lapse of a 1 minute countdown the determined result is
displayed. Press button 3 to convert this into various measurement
3
units (*).
7) Press button 4 to save the result together with date and time in the
4
device.
!
Assessment of the measurement value relative to the ideal range defined by you.
Measurement value for this parameter is outside the value range of this method.
Notes
->
(*) Conversion of the measurement result to mg/l (H O )
->
More notes please see at ID 66 Hydrogen Peroxide
87
0 - 300 mg/l
Name on device: 109-DEWAN50-liq
Possible messages in the line below the measurement value:
"low! / high! / good!":
"Overrange! / underrange!":
If after switching on the device you have already performed the
ZERO (baseline) adjustment for the same type, the following
message is displayed:
-> Measure new ZERO
-> Use last ZERO
4
For the sake of precision select "Use last ZERO" only
if the stored ZERO value is used with the
same cuvette and the same water sample for the test!
Possible messages in the line below the measurement value:
"low! / high! / good!":
"Overrange! / underrange!":
(DW50)
= continue with step 2)
= continue with step 4)
2
2
Liquid
87