PCE -IR 53 Serie Guía Del Operario página 7

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

Lock/Unlock
Prevents settings being changed via a four-digit numerical code.
The default password is 1234.
Change Password
Enter, confirm and save a new four-digit code.
Start/Stop Logging
Manually begins or ends data logging (requires MicroSD Card, available
separately).
If Scheduled Start is enabled in Settings > Data Logging, then logging cannot
be started manually.
To manually start logging, you must first disable Scheduled Start.
Graph
Displays the recent history of the Filtered Temperature and the Sensor
Temperature. To scroll backwards and forwards in time, touch the graph and
drag it. The graph stores the most recent 24 hours of temperature data.
Reset Graph
Clears and restarts the graph.
Real-Time Scrolling View
Returns the graph to the real-time scrolling view, showing the most recent
measurements.
Acknowledge Alarms
Switches the relay outputs for triggered alarms to their normal, untriggered
state. The background of the Temperature View, List View and Graph View
will stay red, and the alarms will not be triggered again, until the alarms are
reset (see "Alarms" below). Alarms can be acknowledged when the display
is locked.
Settings
Access the configuration parameters. Press Apply
to save the settings,
or Exit
to leave the screen without saving.
7
Tabla de contenido
loading

Este manual también es adecuado para:

Pce-ir 53

Tabla de contenido