Sirona VDW 6:1 Instrucciones De Uso página 39

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

Operating instructions VDW 6:1
61 34 378 D3451.201.04.06.02
➢ Sterilize the instrument in the steam sterilizer with saturated water
vapor.
Unit pressure in bar and psi ww
Overpressure:
Temperature:
Holding time:
Steam sterilizers that comply with either EN 13060 class B (for
DAC PROFESSIONAL _ instruments
example, DAC PREMIUM / DAC PROFESSIONAL) or EN 13060
class S and are also suitable for the sterilization of straight/contra-angle
handpieces are approved.
NOTICE! Do not exceed 140°C (284°F), even during the drying phase.
After sterilizing
After sterilizing - heading
1. Remove the instrument from the steam sterilizer immediately.
Subsequent steps_instruments
CAUTION! The instrument is hot. Risk of burns!
2.04 bar (29.59 psi)
134° C (274° F)
3 min
7 Conditioning
37
Tabla de contenido
loading

Tabla de contenido