7 Conditioning
134°C
36
NOTICE! Only use Sirona T1 spray.
Using the T1 spray
Procedure
✔ The instrument is clean and disinfected.
Vorgehensweise _ Winkelstück
1. Press the contra-angle handpiece head with the chuck firmly against
the spray can nozzle.
2. Spray the chuck for 1 - 2 seconds.
IMPORTANT: Hold the spray can upright.
3. Wipe any spray that comes out with a disinfection cloth.
7.5

Sterilizing

✔ The instrument is clean and disinfected.
Action steps_instruments
274°F
✔ Instrument maintenance is complete.
✔ If necessary, the instrument can be sterilized in packaging suitable
for sterilization and storage: paper/plastic composite packaging
according to ISO 11607.
Operating instructions VDW 6:1
61 34 378 D3451.201.04.06.02