Emission - Chauvin Arnoux C.A 61 Manual De Instrucciones

Calibradores de proceso y de temperatura
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

Resolution : 0.1°C from 100 to +300°C for couples K, T, J, L and N
0.5°C from 50 to +300°C for couples R and S
1°C above these values
Series mode rejection (10 mV/50 Hz), K couple: < 0.8°C
Max permitted voltage in common mode: 60 V AC or 85 V peak
Common mode rejection (10 V DC or 50 Hz), K couple: < 0.3°C
Temperature coefficient: £ 10% of the accuracy /°C
Max permitted voltage on the IN terminals: 100 V DC or AC
4.3.2 Emission mode
The measurement current is counted positively when it is a question of the OUT+ terminal
C.A 61
DC voltage emission
Range
12 V
Time for establishment : £ 0.1 s on resistive load
Temperature coefficient: £ 10% of the accuracy /°C
Permitted voltage on the OUT terminals: 30 V DC or AC peak
Internal resistance: £ 1 W . Output current: 20 mA max
Setpoint resolution: 0.1 V
DC current emission
Range
24 mA
(1) If the load is too strong, the OUT symbol flashes on the screen
Time for establishment: £ 0.1 s on resistive load
Temperature coefficient: £ 10% of the accuracy /°C
Max permitted voltage on the OUT terminals: 30 V DC or AC peak
Scale available calibrated in % (0% = 4 mA and 100% = 20 mA)
Setpoint resolution: 0.1 mA
C.A 63
Simulation of thermometric resistance
Sensor
Pt 100 W
-200...850°C (-328...1562°F)
Resolution : 0.1°C from -100 to +300°C (1°C beyond these values)
Nominal measurement current: 1 mA
Permitted measurement current: 0.5...3 mA
Emission range
0.00...12.00 V
Emission range
0.00...24.00 mA
Measurement extent
Accuracy over 1 year
±(0.1% rdg + 10 mV)
Accuracy over 1 year
±(0.1% rdg + 20 µA)
Accuracy over 1 year
±(0.1% rdg + 0.2°C)
47
Load resistance
³ 1000 W
Load resistance
£ 500 W
(1)