Flat Head Screw Driver
5.2
Installing Fuse:
1. Remove the fuses from the power cord kit.
2. Install the appropriate fuse for the voltage source used.
• North American fuse: F 6.3AL, 250 VAC (5 x 20 mm).
• European fuse: F 3.15AL, 250 VAC (5 x 20 mm).
3. U se a small flathead screw driver to depress the latch and open the fuse drawer as shown in Figure 1.
Figure 1: Fuse drawer opening
4. Install a fuse into the rear clip as shown in Figure 2.
5. Install a second fuse into the spare fuse holder located at the front of the drawer as shown in Figure 2 .
6. Close fuse drawer and ensure the drawer is securely latched.
Figure 2: Fuses installed

5. 3 Connecting the Power Cord:

1. Remove the power cord from the power cord kit.
2. Check to ensure that the plug on the power cord is
compatible with the applicable power receptacle in
the region of use. If the power cord is not compatible
contact an Ecolab/Microtek representative.
3. Attach the power cord to the power entry module
located below the main power switch on the
back of the IntraTemp Solution Warmer.
4. Remove the power cord retention bracket and two mounting screws from the power cord kit.
5. Using a #1 Phillips Head Screw Driver and furnished screws connect the power cord
retention bracket to the IntraTemp Solution Warmer via the threaded connectors
located to each side of the power inlet module (see Figure 3).
9 of 128
Figure 3: Power Cord Installation