• North American fuse: F 6.3AL, 250 VAC (5 x 20 mm).
• European fuse: F 3.15AL, 250 VAC (5 x 20 mm).
• Temperature Scale: Celsius or Fahrenheit selectable by qualified service personnel.
• Basin Temperature Set Point: selectable by qualified service personnel
in a range from 37 to 43 ± 1°C (98 to 110 ± 1.5°F).
• Heated Bottle Pocket Receptacles Temperature Set-Point: Factory setting
per Customer Specification up to 43 ± 1°C (110 ± 2°F).
• Use only with Ecolab ThermaBasin; product code ITB100.
• Minimum fluid volume: 1.5 Liters.
• Equipment mass including working load is 43 Kg
• EMC Compliance: The ITW6L has been evaluated with respect to electric shock, fire and mechanical
hazards in accordance with EN IEC 60601-1, AAMI ES60601-1, CAN/CSA C22.2 – No. 60601-1.
Guidance and manufacturer's declaration – electromagnetic emissions
The ITW6L is intended for use in the electromagnetic environment specified below. The customer or the user of
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 6100-3-2
Voltage fluctuations
flicker emissions
IEC 6100-3-3
Guidance and manufacturer's declaration – electromagnetic immunity
The ITW6L is intended for use in the electromagnetic environment specified below. The customer or the user of
Immunity test
Electrostatic
±6 kV contact
discharge (ESD)
IEC 6100-4-2
±8 kV air
Electrical fast
±2 kV for power
transient/burst
supply lines
IEC 6100-4-4
±2 kV for input/output
lines
Surge
±1 kV differential
IEC 4100-4-5
mode
±2 kV common mode
the ITW6L should assure that it is used in such an environment.
Compliance
Group 1
Class B
Class B
Complies
the ITW6L should assure that it is used in such an environment.
IEC 60601 test
Compliance level
level
±6 kV contact
±8 kV air
±2 kV for power
supply lines
Not applicable
±1 kV differential
mode
±2 kV common mode
Electromagnetic environment – guidance
The ITW6L uses RF energy only for its internal function. Therefore, its
RF emissions are very low and are not likely to cause any interference in
nearby electronic equipment.
The ITW6L is suitable for use in all establishments, including domestic
establishments and those directly connected to the public low-voltage
power supply network that supplies buildings used for domestic purposes.
Electromagnetic environment – guidance
Floors should be wood, concrete, or ceramic tile. If
floors are covered with synthetic material, the
relative humidity should be at least 30%.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment.
5 of 128