5 - Basic Set-up Examples
Requirement
Change the Output Phase
The soft-key labelled Phase will be selected as default.
Enter a phase of -45 degree.
•
The set phase angle is the point in the waveform period which is coincident with the Sync or trigger
edge, i.e. it is the point in the period at which the waveform starts. Hence a negative phase setting
advances, and a positive phase setting delays the waveform relative to the Sync or trigger; the
waveform in the graph box changes to show this.
The Align soft-key is used to re-align phase when making frequency changes.
Change the Output Polarity
Press the soft-key labelled Type to invert the output polarity.
•
Successive presses of the type key alternates between normal and inverted.
29
TGF4000 Series Quick Start Guide