Ref Clock Output; Ref Clock Input; Phase Synchronising Two Generators - Aim TTi TG5011A Manual De Instrucciones

Tabla de contenido
Sync Outs (Contd)
Sweep Sync:
Output Signal
Level:

Ref Clock Output

Buffered version of the 10MHz clock currently in use (internal or external)
Output Level:
Inputs
Trig In
Frequency Range:
Signal Range:
Minimum Pulse Width:
Polarity:
Input Impedance:
External Modulation Input (for AM, FM, PM, PWM)
Voltage Range:
Input Impedance:
Bandwidth:

Ref Clock Input

Input for an external 10MHz reference clock
Voltage Range:
Maximum Voltage:
Minimum Voltage:

Phase Synchronising Two Generators

Two generators can be synchronised together to provide outputs at the same frequency (or
harmonics) and with a phase difference. The amplitude and phase of these outputs can also be
modulated providing the capability to perform QAM and QPSK respectively. It is also possible to
synchronise more than two generators but results are not guaranteed.
Carrier Waveforms:
Phase:
(exept Pulse
& PRBS)
Marker Off
A square wave that is a TTL high from the beginning of the
sweep and a TTL low from the midpoint of the sweep
Marker On
A square wave that is a TTL high from the beginning of the
sweep and a TTL low from the marker frequency
Logic level nominally 3V.
Nominally 3V logic level from 50Ω.
DC − 1MHz.
Threshold nominally TTL level; maximum input ±10V.
50ns
Selectable as high/rising edge or low/falling edge.
10kΩ
± 5V full scale
5kΩ typical
DC to 20kHz
1Vpp – 5Vpp
+5V
-1V
Sine, Square, Ramp, Arb, Pulse, PRBS
Range
Resolution
Accuracy
-360.0 to +360.0 degrees
0.1 degree
< ±5ns