COmPRESSOR CLEANING

1.
With power switch in the "Off" position, unplug power cord from wall outlet.
2.
Wipe outside of the compressor cabinet with a clean damp cloth every few days to keep dust-free.
DANGER
Do not submerge in water; doing so will result in compressor damage.

FILTER CHANGE

1.
Filter should be changed every 6 months or sooner if filter discolors.
2.
Remove filter (Fig. A) by grasping it firmly and pulling out of the unit. Discard old filter.
3.
Replace with a new DeVilbiss filter (part #3655D-601). Additional filters should be purchased from your DeVilbiss
provider.
4.
Push new filter into position.
CAUTION– Do not reuse or wash filter or substitute any other material such as cotton for a DeVilbiss air-inlet filter. It will
result in compressor damage.

mAINTENANCE

All maintenance must be performed by a qualified DeVilbiss provider or authorized service center.
DANGER
Electric shock hazard. Do not remove compressor cabinet. All disassembly and maintenance must be done by a
qualified DeVilbiss provider.

SPECIFICATIONS

Size. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.5" W x 4.0" H x 7.2" D (19.1 cm x 10.2 cm x 18.3 cm)
Weight
3655D . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3.8 lbs. (1.72 kg)
3655I . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4.8 lbs. (2.18 kg)
Electrical Requirements
3655D . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.5 amps, 115 VAC (±10%) 60 Hz
3655I . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.4 amps, 230 VAC (±10%) 50 Hz
Sound Level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58-62 dBA
Power Consumption
3655D . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 140 watts maximum
3655I . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 160 watts maximum
Maximum Compressor Pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35 psig (241 kPa)*
Compressor Free Air Flow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 lpm*
Operating Temperature Range. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . +40° to +104°F (+5 to +40°C)
Operating Humidity. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .up to 95% non-condensing
Storage/Transport Temperature Range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . -40° to +158°F (-40° to +70°C)
Storage/Transport Humidity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .up to 95% non-condensing
Supplied Nebulizer Capacity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 ml (cc)
Supplied Nebulizer Nebulization Rate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0.15 ml/min or greater*
Supplied Nebulizer MMAD . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 microns or less**
Supplied Nebulizer Operating Pressure. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12 psig (min)*
*Conditions may vary based on altitude above sea level, barometric pressure, and temperature.
**Value determined with minimum flow rate through the nebulizer (20°C, sea level).
There are no hazardous materials used in the construction of this unit, therefore, there are no known risks associated with
disposal of the equipment.