Maintenance; Clean The Instrument; Store The Instrument; Replace The Batteries - Hach sensION+ Serie Manual Del Usuario

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 40
5. Select a folder location and give the data set a file name.
6. Select the type of file (.txt or .csv), then click Save.
Figure 11 Send data to a computer

Maintenance

Clean the instrument

N O T I C E
Never use cleaning agents such as turpentine, acetone or similar products to
clean the instrument including the display and accessories.
N O T I C E
Make sure that all meter connectors stay dry. Use a dry towel to remove liquid
from the connectors.
Clean the exterior of the instrument with a moist cloth and a mild soap
solution.

Store the instrument

W A R N I N G
Explosion hazard. Expired batteries can cause hydrogen gas buildup
inside the instrument. Replace the batteries before they expire and do
not store the instrument for long periods with the batteries installed.
N O T I C E
To avoid potential damage to the meter from battery leakage, remove the meter
batteries prior to extended periods of non-use.

Replace the batteries

W A R N I N G
Explosion hazard. Incorrect battery installation can cause the release of explosive
gases. Be sure that the batteries are of the same approved chemical type and
are inserted in the correct orientation. Do not mix new and used batteries.
English 15

Hide quick links:

Capítulos

Tabla de contenido
loading

Tabla de contenido