Field Of Application; Recommended Sterilisation Method - HEBU medical HE 3110-11 Instrucciones Para El Uso

Idiomas disponibles

Idiomas disponibles

EN
HE3530GR
1/2 spare lid alone without filter, green, 285x280mm
HE3530RO
1/2 spare lid alone without filter, rot, 285x280mm
HE3530SI
1/2 spare lid alone without filter, silver, 285x280mm
HE3530SW
1/2 spare lid alone without filter, black, 285x280mm
HE3510BL
1/1 spare lid alone without filter, blue, 580x280mm
HE3510GO
1/1 spare lid alone without filter, gold, 580x280mm
HE3510GR
1/1 spare lid alone without filter, green, 580x280mm
HE3510RO
1/1 spare lid alone without filter, rot, 580x280mm
HE3510SI
1/1 spare lid alone without filter, silver, 580x280mm
HE3510SW
1/1 spare lid alone without filter, black, 580x280mm

4 Field of application

The HEBUmedical Sterilization Container System is a sterile packaging system for medical
instruments and textiles that utilizes an established filter technology. It is a reusable device
and features an assortment of sizes and configurations that provide an effective packaging
method for sterilization, storage and transportation of surgical instrument by healthcare
providers. This container system is intended for use in pre-vacuum Steam Sterilizers.
It is intended to allow sterilisation of the enclosed device and also maintain sterility of the
enclosed device until used.
All Containers are available with perforated and non-perforated bottom and perforated lids.
All containers use suitable disposable paper filters, which must be replaced before each
sterilisation.

4.1 Recommended sterilisation method

In order to ensure sterile safety, tests were carried out by an independent and accredited test
laboratory. The purpose of these trials was to validate a sterilisation process for the reusable
HEBUmedical sterilisation container with steam.
The HEBUmedical Container System was validated with the following sterilisation
parameters. Each facility may need to run internal testing to determine if adjustments are
necessary for their facility.
method:
three times pre-vacuum steam sterilization
temperature:
132 °C (270 °F)
holding time:
4 minutes
Drying time:
10 minutes
loading:
standard surgical instruments (scissors, clamps, forceps) and textiles
The validation considered the following Containers of the HEBUmedical Container System:
Full-size container,
¾-size container,
½-size container,
20
Container systems
loading