2.2 Pneumatic Pressure (Figure 2)
1. Refer to the Data sheet for the correct pressure mediums to be used.
2. Connect the Unit Under Test (UUT) to the required connection port.
Note: For instruments with NPT connections, use applicable bonded sealing as shown in figure
below.
ISO 228 G1/8
bonded seal
Recommended method

2.3 Pneumatic connections

WARNINGS
PARALLEL THREADS MUST BE USED. FEMALE THREAD TYPE IS PARALLEL THREAD TO
ISO228/1 (DIN ISO228/1, JIS B0202) G1/8.
TAPERED THREADS NOT ALLOWED.
PACE rear
panel
PACE rear
panel
Connection
Input
Reference
For examples of adaptors (see Page 8).
K0467 Revision A
Figure 2 Sealing Pneumatic Connections
PACE pressure port
PACE pressure port
ISO228/1 G 1/8 parallel threads (DIN ISO228/1, JIS B0202)
ISO228/1 G 1/8 parallel threads (DIN ISO228/1, JIS B0202)
ISO 228 G1/8
bonded seal
Alternative method below 100 bar
Parallel threads must be used.
Female thread type is parallel thread to
ISO228/1 (DIN ISO228/1, JIS B0202)
G1/8
Tapered threads NOT allowed.